DDG-2519 Industrial Online Conductivity Instrument
Features:
Equipped with the latest single chip microcomputer, this conductivity meter is characterized by its |
low power consumption, less peripheral components and high performance, stability and reliability. |
It adopts the LCD display. When you push the keys, the conductivity meter can display the |
conductivity and temperature values respectively. |
Three types of electrode constants can be used in this meter. |
|
The four levels of measuring ranges can shift automatically. |
It is available with automatic temperature compensation. |
4~20mA isolated current output. |
Technical Indexes
1. Working conditions |
a. Ambient temperature: 0~55℃ |
b. Relative humidity:<85% |
c. Power supply: 24 VDC±5% |
d. Solution temperature tested: 5~55℃ |
2. Conductivity measuring range: K=0.01 electrodes: 0~20.0 µS/ cm |
K=0.1 electrodes: 0~200.0 µS/ cm |
K=1.0 electrodes: 0~2000 µS/ cm |
Temperature measuring range: 0-100℃ |
3. Intrinsic error of conductivity measurement: ±l%F.S |
Intrinsic error of temperature measurement: ±0.5℃ |
4. Temperature compensation range and influence error: 5~55℃, ±0.5%F.S |
5. Isolated current output and intrinsic error: 4~20mA (RL ≤500Ω) ± 0.5%F.S |
6. The installation type of electrode: Pipeline type and circulation type. |