Item No.: 266761
OverviewQuick DetailsPlace of Origin:Shanghai, China (Mainland)Brand Name:BOQUModel Number:YLG-2058Resolution:0.01mg/LTemperature:0 ~ 60.0
INQUIRY
Description
Overview
Quick Details
Place of Origin:
Shanghai, China (Mainland)
Brand Name:
BOQU
Model Number:
YLG-2058
Resolution:
0.01mg/L
Temperature:
0 ~ 60.0 ℃
Packaging & Delivery
Packaging Details
carton box
Delivery Time
5-10 working days

YLG-2058 Industrial Online Residual Chlorine Tester

 YLG-2058 Industrial o<em></em>nline Residual Chlorine Tester

 

YLG2058 Industrial Online Residual Chlorine Analyzer is a brannew residual chlorine analyzer in our company; It is a 

highintelligence online monitor, It is made up of three parts: a secondary instrument and a sensor, a organic glass flow 

cell. It can measure residual chlorine, pH and temperature simultaneously. It can be widely used for continuous monitoring 

of residual chlorine and pH value of various water quality in power, water plants, hospitals and other industries.

 

 

 Feateurs:

English Display, English Menu operation: Easy operation, English prompts during the whole operational 

procedure ,convenient and quick.

Intelligent: It adopts high-precision AD conversion and single chip microcomputer processing technologies and

can be used for the measurement of the PH values and temperature, automatic temperature compensation and

self-checking etc function.

Multi-parameter display: On the same screen, residual chlorine, temperature, pH value, output current, status

and time are displayed.

Isolated current output: Optoelectronic isolating technology is adopted. This meter has strong interference

immunity and the capacity of long-distance transmission.

High and low alarm function: High and low alarm isolated output, hysteresis can be adjusted

 

 

Technical index: 

1. Measuring range     

Residual chlorine: 0-20.00mg/L,      Resolution: 0.01mg/L;

HOCL: 0-10.00mg/L       Resolution: 0.01mg/L;

pH value: 0 - 14.00pH        Resolution: 0.01pH;

Temperature: 0- 99.9 ℃          Resolution: 0.1 

2. Accuracy: Residual chlorine: ± 2% or ± 0.035mg / L, take the bigger;

HOCL: ± 2% or ± 0.035mg / L, take the bigger;

pH value: ± 0.05Ph

Temperature: ± 0.5 ℃ (0 ~ 60.0 ℃);

3. Sample temperature: 0 ~ 60.0 ℃, 0.6MPa;

4. Sample flow rate: 200 ~250 mL/1min  Automatic and Adjustable

5. Minimum detection limit: 0.01mg / L

6. Isolated current output: 4~20 mA(load <750Ω)

7. High and low alarm relays: AC220V, 7A; 

hysteresis 0- 5.00mg / L, arbitrary regulation

8. RS485 communication interface (optional)

It can be convenient to computer monitoring and communication

9. The data storage capacity: 1 month (1 point/5 minutes)

10. Power Supply: AC220V ± 22V, 50Hz ± 1Hz; DC24V (optional).

11. Protection grade: IP65

12. Overall dimension: 146 (length) x 146 (width) x 108 (depth) mm;

dimension of the hole: 138 x 138mm

Note: The wall installation may be ok, please specify when ordering.

13. Weight: Secondary Instrument: 0.8kg, flow cell with residual chlorine, 

pH electrode weight: 2.5kg;

14. Working Conditions: ambient temperature: 0 ~ 60 ℃;  relative humidity <85%; 

15.Adopt the flow-through installation, inlet and outlet diameter at Φ10. 

 